• Pressure ranges to 36,000 psi (2500 bar)
• Hart Communication capability
• Measure mA or V, and with 24V loop power
• Easy-to-use, inexpensive pressure calibrator with uncertainty better than 0.025%FS
With advanced microprocessor technology and state-of-the-art silicon pressure sensors, the ConST 273 series precision pressure calibrators provide a pressure calibration solution for gauges, transmitters, and switches in a wide pressure range. In a highly portable package, it can not only measure pressure precisely with a built-in pressure sensor, but can also read the current or mV produced by the transducer, and can even supply an excitation voltage to power sensors or transmitters during calibration. To reach 0.025% F. S. accuracy over ranges to 10,000 psi (700 bar) and 0.1% F. S. accuracy over range to 36,000 psi (2,500 bar), every silicon pressure sensor in gauges has been specially aged, tested and screened before assembly. ConST 273 series precision pressure calibrators are unmatched in performance and reliability.
• Pressure measurement uncertainty:
ConST 273-02: 0.025%FS
ConST 273-05: 0.05%FS
• Pressure ranges to 36,000 psi (2500 bar)
• Measure mA or V with 0.02% RD + 0.003%FS accuracy
• Power transmitters during test using 24V loop supply
• Pressure switch test
• Hart Communication capability
• Advanced temperature compensation
• Dual readout
• Min/Max/Hold to capture changing measurements
• Data logging
• Large, easy to read display with 6-digit resolution
• Back lighted display • Rechargeable battery or AC adapter
• NIM traceable calibration (optional)
• Gauge calibration
• Precision pressure measurement
• Pressure transmitters calibration
• Pressure switch testing
• Safety relief valve resting
• Pressure regulator testing
• Hydraulic leak testing
• Accuracy
ConST 273-02: 0.025% of full scale
ConST 273-05: 0.05% of full scale
• Gauge types
Gauge pressure
Compound Pressure
Absolute pressure
Differential pressure
• Pressure Ranges
|
Gauge Pressure [1]
|
|||||
|
P/N
|
Pressure
(psi)
|
Pressure
(bar)
|
Media [2]
|
Accuracy
(% FS)
|
Burst Pressure
|
|
V15
|
-15
|
-1.0
|
G
|
0.025 (0.05)
|
3 X
|
|
GP2
|
2
|
0.16
|
G
|
0.05
|
3 X
|
|
GP5
|
5
|
0.35
|
G
|
0.025 (0.05)
|
3 X
|
|
GP10
|
10
|
0.7
|
G
|
0.025 (0.05)
|
3 X
|
|
GP15
|
15
|
1.0
|
G , L [3]
|
0.025 (0.05)
|
3 X
|
|
GP30
|
30
|
2.0
|
G , L [3]
|
0.025 (0.05)
|
3 X
|
|
GP50
|
50
|
3.5
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP100
|
100
|
7.0
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP300
|
300
|
20
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP500
|
500
|
35
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP600
|
600
|
40
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP1K
|
1000
|
70
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP2K
|
2000
|
140
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP3K
|
3000
|
200
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP5K
|
5000
|
350
|
G , L
|
0.025 (0.05)
|
3 X
|
|
GP10K
|
10000
|
700
|
G , L
|
0.025 (0.05)
|
2 X
|
|
GP15K
|
15000
|
1000
|
G , L
|
0.1
|
2 X
|
|
GP25K
|
25000
|
1600
|
G , L
|
0.1
|
1.5 X
|
|
GP30K
|
30000
|
2000
|
G , L
|
0.1
|
1.5 X
|
|
GP36K
|
36000
|
2500
|
G , L
|
0.1
|
1.5 X
|
|
Absolute Pressure
|
|||||
|
P/N
|
Pressure
(psi)
|
Pressure
(bar)
|
Media
|
Accuracy
(% FS)
|
Burst Pressure
|
|
AP5
|
5
|
0.35
|
G
|
0.1 (0.2)
|
3 X
|
|
AP10
|
10
|
0.7
|
G
|
0.1 (0.2)
|
3 X
|
|
AP15
|
15
|
1.0
|
G
|
0.1 (0.2)
|
3 X
|
|
AP30
|
30
|
2.0
|
G
|
0.1 (0.2)
|
3 X
|
|
AP50
|
50
|
3.5
|
G
|
0.1 (0.2)
|
3 X
|
|
AP100
|
100
|
7.0
|
G , L
|
0.1 (0.2)
|
3 X
|
|
AP300
|
300
|
20
|
G , L
|
0.1 (0.2)
|
3 X
|
|
AP500
|
500
|
35
|
G , L
|
0.1 (0.2)
|
3 X
|
|
AP1K
|
1000
|
70
|
G , L
|
0.1 (0.2)
|
3 X
|
|
AP3K
|
3000
|
200
|
G , L
|
0.1 (0.2)
|
3 X
|
|
AP5K
|
5000
|
350
|
G , L
|
0.1 (0.2)
|
3 X
|
|
Differential Pressure
|
|||||
|
P/N
|
Pressure
(inH2O)
|
Pressure
(mbar)
|
Media
|
Accuracy
(% FS)
|
Burst Pressure
|
|
DP1
|
1
|
2.5
|
G
|
0.4
|
100 X
|
|
DP2
|
2
|
6.0
|
G
|
0.2
|
100 X
|
|
DP5
|
5
|
10
|
G
|
0.1
|
50 X
|
|
DP10
|
10
|
25
|
G
|
0.1
|
20 X
|
|
DP20
|
20
|
50
|
G
|
0.05
|
20 X
|
|
DP30
|
30
|
100
|
G
|
0.05
|
20 X
|
|
Compound Pressure
|
|||||
|
P/N
|
Pressure
(psi)
|
Pressure
(bar)
|
Media
|
Accuracy
(%FS)
|
Burst Pressure
|
|
CP2
|
±2
|
±0.16
|
G
|
0.05
|
3X
|
|
CP5
|
±5
|
±0.35
|
G
|
0.025 (0.05)
|
3X
|
|
CP10
|
±10
|
±0.7
|
G
|
0.025(0.05)
|
3X
|
|
CP15
|
±15
|
±1
|
G
|
0.025(0.05)
|
3X
|
|
CP30
|
-15 to 30
|
-1 to 2
|
G
|
0.025(0.05)
|
3X
|
|
CP1K
|
-15 to 1000
|
-1 to 70
|
G
|
0.025(0.05)
|
3X
|
|
Electrical Measurement and Source Accuracy
|
||
|
|
Range
|
Accuracy
|
|
Voltage DC
|
±30.0000V
|
± (0.02%RD+0.003%FS)
|
|
Current DC
|
±30.0000mA
|
± (0.02%RD+0.003%FS)
|
|
DC 24V
|
24V±0.5V max:50mA Protect at: 120mA
|
|
|
Switch [1]
|
Status OPEN/CLOSED
|
|
Note:
[1] 1~12V if switch has detective voltage
• Display
Description: Dual-line 6 full digit FSTN LCD with LED Backlight
Display rate: 3.5 readings per second (Default setting)
Numeral display height: 16.5mm (0.65”)
• Pressure Units
Pa, kPa, MPa, psi, bar, mbar, inH2O, mmH2O, inHg, mmHg,kgf/cm2
• Environmental
Compensated Temperature: 14°F to 122°F (-10°C to 50°C)
Operating Temperature*: 14°F to 122°F (-10°C to 50°C)
*0.025%FS accuracy guaranteed only over the ambient temperature range of 68°F to 79°F (20°C to 26°C) Storage Temperature: -4°F to 158°F (-20°C to 70°C)
Humidity: <95%
• Media Compatibility
(0 – 30) psi: Clean dry non-corrosive gases
(50 – 36,000) psi: media compatible with 316L SS
• Pressure Port
1/4”NPT(≤15,000psi)
Autoclave F-250-C, 9/16" - 18 UNF female or Autoclave M-250-C, 9/16"-18 UNF male(>15,000psi)
0.156 inch (4mm) test hose (for differential pressure)
Other connections are available per request
• Electrical Connection: 0.156 inch (4mm) sockets
• Over Pressure Warning: 120%
• Data Logger
Storage capacity: 30 files, 40 records per file
Mode: manual and automatic
Hourly-record: record the data every hour
Interval-record: set by user
• Power Battery:
Recharge Li-ion polymer battery
Li-Battery working time: 40 hours
Rechargeable time: 4 hours
External power: 110V/220V power adapter(10V DC)
• Enclosure
Case material: Aluminum alloy
Wetted parts: 316L SS
Dimension: 120mm diameter X 46mm depth X 184mm height
Weight: 0.7kg
• Communication
RS232 (DB9/F, environmentally sealed)
Baud rate: 1200,2400,4800,9600
Data length: 8 bits
Stop bit: 2 bits
Address: from 1 to 112
• Accessories (included)
110V/220V external power adapter (DC 10V)
2 pieces 1.5-meter test leads
2 pieces 0.156 inch (Ø4 mm) test hose (for differential pressure gauge only)
Ordering Information
Model number: ConST273

Accessories (Included)
1. Rechargeable Li-ion polymer battery 1 pc
2. 110V/220V external power adapter (DC10V) 1 pc
(Available standards: USA, EU)
3. 1.5-meter test leads 2 pcs
4. Short circuit cable 2 pcs
5. 0.156 inch (Ø4 mm) test hose(for DP gauge only) 2 pcs
6. User’s manual 1 pc
ConST/land software (free download at www.constek.com)
Optional Accessories
|
Model Number
|
Description
|
|
9702
|
Spare rechargeable Li-ion polymer battery for ConST273
|
|
9814
|
Spare 110V/220V external power adapter (DC 10V)
(USA and EU standard are available)
|
|
9902
|
Carrying case for ConST273 (up to 2pcs)
|
|
9050
|
USB to RS 232 (DB-9 Male) Adapter
|
|
9502
|