• Sourcing, simulating and measuring pressure, temperature and electrical signals
• Smartphone-like menu and interface make the operation simpler and easier
• The internal cold junction compensation sensor can be re-calibrated easily
• Ultra-compact, 3.9” x 7.6” x 2.0”, and 1.6 lb (0.7 kg)
A highly integrated multifunction calibrator featuring several patented technologies, ConST 317A is an ultra-compact, rugged, and ease-to-use hand-held device for sourcing, simulating and measuring pressure, temperature, and electrical signals. Its smartphone-like menu and interface make the operation simpler and easier. ConST 317A is ideal for calibrating, maintaining, and troubleshooting process instrumentation. Automation and documentation capabilities make the ConST 317A a turnkey solution.
Sourcing, simulating and measuring pressure, temperature and electrical signals
• Sources and measures mV, mA, ohms, RTDs, thermocouples, frequency, and pulses
• Simulates and measures 13 thermocouples, 11 RTDs, and pressure to calibrate transmitters
• Measures and sources pressure using CDP series Intelligent Digital Pressure Modules from -15 psi to 10,000 psi (-1 bar to 700 bar)
• 24V loop power supply
• Simultaneous dual reading capability
• Automatic switch test
• Supports square root transmitter
• Pulse frequency output for the calibration of flow totalizer
Easy to use
• Smartphone-like menu and interface make the operation simpler and easier
• Ultra-compact, size 3.9” x 7.6” x 2.0” (100mm x 192mm x52mm), and weight 1.6 lb (0.7 kg)
• One hand operation
Calibrated cold junction compensation (Patented)
• Cold junction equivalent block in the calibrator
• A calibrated PRT element with flexible leads is installed in the equivalent block for thermocouple cold junction compensation
• This PRT element can be pulled out from the calibrator and re-calibrated and corrected at ice point easily
Built-in temperature readout
CVD coefficients of a calibrated PRT can be input into the calibrator for accurate temperature measurement.
Documenting and automated procedure capability
• Manage the information of the device under test
• Set up automated calibration procedures, and ConST317A performs the test, calculates the errors, displays and/or stores the results in the memory, and highlights the tolerance points.
• As-found and As-left functions allow recording and documenting results for quality control.
• Download tasks and upload the results
• Snapshots allow you to capture and save work
Build-in unit conversion tool
Build-in converters for pressure units, temperature units, temperature vs. resistance (RTDs), and temperature vs. millivolt (thermocouples)
Multi lingual Interface
English, German, French, Italian, Spanish, Portuguese, Simplified Chinese
(Traditional Chinese, Japanese and Russian are available per request)
Display
3.5 inch TFT color screen
Rugged
• Rugged design for harsh environments.
• Passed a 1-meter drop test
• Three years warranty for the ConST 317A, and one year for the battery pack
Misuse protection
Up to 30V voltage on any two sockets and up to 1A current on current sockets will not damage the calibrator, and the calibrator will return to normal condition as soon as the voltage or current is removed.
Rechargeable and replaceable battery
• Rechargeable Li-ion battery for 15 hours uninterrupted use.
• Battery life will be reduced when 24V is applied.
• The rechargeable battery is replaceable.
The ConST 317A multifunction calibrator is a process tool for measuring, sourcing and simulating mA, mV, V, RTDs, thermocouples, ohms, frequency, pulses and pressure, captures switch values and provides 24 V loop power.
• Electrical
Resistance measurement / simulation
Voltage measurement / generation
Current measurement /generation
Frequency measurement / generation
Pulse counting/generation
24 VDC loop supply
Switch sensing
• Temperature
RTD measurement /simulation
TC measurement /simulation
• Pressure
Gauge pressure measurement
Absolute pressure measurement
Electrical Specifications
| Measurement accuracy | ||||
| Range | Resolution | Accuracy | ||
| Voltage DC | -75.0000 to 75.0000 mV | 0.1µV | 0.01%RD+0.005%FS | |
| -30.0000 to 30.0000 V | 0.1mV | 0.01%RD+0.005%FS | ||
| Current DC | -30.0000 to 30.0000 mA | 0.1µA | 0.01%RD+0.005%FS | |
| Resistance | Two-wire | 0 to 400.000 ohm | 1 mΩ | 0.02%RD+0.005%FS |
|
Three-wire
|
0 to 400.000 ohm | 1 mΩ | 0.02%RD+0.005%FS | |
|
Four-wire
|
0 to 400.000 ohm | 1 mΩ | 0.01%RD+0.005%FS | |
|
Two-wire
|
0 to 4000.00 ohm | 10 mΩ | 0.02%RD+0.005%FS | |
|
Three-wire
|
0 to 4000.00 ohm | 10 mΩ | 0.02%RD+0.005%FS | |
|
Four-wire
|
0 to 4000.00 ohm | 10 mΩ | 0.01%RD+0.005%FS | |
| Frequency | 1 to 50000.0 Hz | 0.1Hz | 0.005%RD+0.002%FS | |
| Pulse | 0 to 999999 | 1 | N/A | |
| Limit Switch | If the switch has detective voltage, its range is +3v to +24V | |||
| Source accuracy | ||||
| Voltage DC | -10.000 to 75.000 mV | 1µV | 0.02%RD+0.005%FS | |
| 0 to 12.0000 V | 0.1mV | 0.02%RD+0.005%FS | ||
| Current DC | 0 to 22.000 mA | 1µA | 0.02%RD+0.005%FS | |
| Resistance | 1 to 400.00 ohm | 10 mΩ | 0.02%RD+0.005%FS | |
| 1 to 4000.0 ohm | 100 mΩ | 0.03%RD+0.01%FS | ||
| Frequency | 0 to 50000.0 Hz | 0.1Hz | 0.005%RD+0.002%FS | |
| Pulse | 0 to 999999 | 1 | N/A | |
| DC24V | N/A | N/A | 0.5V | |
Temperature Specifications
| Thermocouple Measurement and Source accuracy | |||||
| Measure and Simulate | Standard | Temperature Range (°C) | Accuracy (°C) | ||
| Measure | Source | ||||
| S | IEC 584 | -50 to 1768 | -50 to400 | 1.0 | 1.1 |
| 400 to 1000 | 0.6 | 0.6 | |||
| 1000 to1768 | 0.7 | 0.8 | |||
| R | IEC 584 | -50 to 1768 | -50 to 200 | 1.4 | 1.4 |
| 200 to 500 | 0.6 | 0.6 | |||
| 500 to 1768 | 0.6 | 0.7 | |||
| B | IEC 584 | 0 to 1820 | 50 to 450 | 3.8 | 3.8 |
| 450 to 800 | 0.9 | 0.9 | |||
| 800 to 1820 | 0.7 | 0.7 | |||
| K | IEC 584 | -270 to 1372 | -250 to -200 | 1.0 | 1.1 |
| -200 to -100 | 0.4 | 0.5 | |||
| -100 to 600 | 0.3 | 0.3 | |||
| 600 to 1372 | 0.4 | 0.5 | |||
| N | IEC 584 | -270 to 1300 | -250 to -200 | 1.5 | 1.6 |
| -200 to -100 | 0.5 | 0.6 | |||
| -100 to 1300 | 0.4 | 0.5 | |||
| E | IEC 584 | -270 to 1000 | -250 to -200 | 0.6 | 0.7 |
| -200 to -100 | 0.3 | 0.3 | |||
| -100 to 0 | 0.2 | 0.2 | |||
| 0 to 700 | 0.2 | 0.3 | |||
| 700 to 1000 | 0.2 | 0.4 | |||
| J | IEC 584 | -270 to 1200 | -210 to -100 | 0.3 | 0.3 |
| -100 to 1200 | 0.3 | 0.4 | |||
| T | IEC 584 | -270 to 400 | -250 to -200 | 0.8 | 0.9 |
| -200 to 0 | 0.4 | 0.4 | |||
| 0 to 400 | 0.2 | 0.2 | |||
| C | ASTM E988 | 0 to 2315 | 0 to 1000 | 0.5 | 0.5 |
| 1000 to 1800 | 0.7 | 0.9 | |||
| 1800 to 2315 | 1.0 | 1.4 | |||
| D | ASTM E988 | 0 to 2320 | 0 to 100 | 0.5 | 0.5 |
| 100 to 1100 | 0.4 | 0.5 | |||
| 1100 to 2000 | 0.6 | 0.9 | |||
| 2000 to 2320 | 0.9 | 1.3 | |||
| G | ASTM E1751 | 0 to 2315 | 0 to 200 | 2.4 | 2.4 |
| 200 to 400 | 0.5 | 0.5 | |||
| 400 to 1400 | 0.4 | 0.5 | |||
| 1400 to 2315 | 0.7 | 1.0 | |||
| L | DIN43710 | -200 to 900 | -200 to -100 | 0.2 | 0.3 |
| -100 to 400 | 0.2 | 0.2 | |||
| 400 to 900 | 0.2 | 0.3 | |||
| U | DIN43710 | -200 to 600 | -200 to 0 | 0.4 | 0.4 |
| 0 to 600 | 0.2 | 0.3 | |||
| RTD Measurement and Source accuracy | ||||||
| Measure and Simulate | Standard | Temperature Range (°C) | Accuracy (°C) | |||
| Measure (2W/3W) | Measure (4W) | Source | ||||
| Pt10(385) | IEC 751 | -200 to 850 | -100 to 200 | 0.65 | 0.60 | 0.65 |
| 200 to 600 | 0.82 | 0.72 | 0.82 | |||
| 600 to 850 | 0.96 | 0.82 | 0.96 | |||
| PT100 (385) | IEC 751 | -200 to 850 | -100 to 200 | 0.15 | 0.1 | 0.15 |
| 200 to 600 | 0.26 | 0.16 | 0.26 | |||
| 600 to 1850 | 0.33 | 0.20 | 0.33 | |||
| Pt100(391) | IEC 751 | -200 to 850 | -100 to 200 | 0.15 | 0.10 | 0.15 |
| 200 to 600 | 0.26 | 0.16 | 0.26 | |||
| 600 to 850 | 0.33 | 0.20 | 0.33 | |||
| Pt100(392) | IEC 751 | -200 to 850 | -100 to 200 | 0.37 | 0.32 | 0.69 |
| 200 to 600 | 0.51 | 0.41 | 0.93 | |||
| 600 to 850 | 0.61 | 0.48 | 1.08 | |||
| Pt500(385) | IEC 751 | -200 to 850 | -100 to 200 | 0.20 | 0.16 | 0.36 |
| 200 to 600 | 0.32 | 0.22 | 0.54 | |||
| 600 to 850 | 0.40 | 0.27 | 0.67 | |||
| Pt1000(385) | IEC 751 | -200 to 850 | -100 to 200 | 0.10 | 0.05 | 0.25 |
| 200 to 600 | 0.2 | 0.10 | 0.42 | |||
| 600 to 850 | 0.27 | 0.14 | 0.54 | |||
| Cu10(427) | IEC 751 | -100 to260 | -100 to260 | 0.61 | 0.56 | 0.61 |
| Cu50(428) | IEC 751 | -50 to 150 | -50 to 150 | 0.17 | 0.13 | 0.17 |
| Cu100(428) | IEC 751 | -50 to 150 | -50 to150 | 0.12 | 0.09 | 0.12 |
| Ni120(672) | DIN 43760 | -100 to260 | -100 to260 | 0.07 | 0.05 | 0.07 |
| Ni100(618) | DIN 43760 | -100 to260 | -100 to260 | 0.08 | 0.06 | 0.08 |
Pressure Specification
| CDP series Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.025% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detail specification refer to pressure modules datasheet. |
General Specifications
| Environmental Specifications | |
| Operating Temperature | -10°C to 50°C |
| Storage Temperature | -20°C to 60°C |
| Humidity | <90%, non-condensing |
| Safety specifications | |
| European Compliance | CE Marked |
| Mechanical Specifications | |
| Display | 3.5 inch TFT color screen |
| Electrical Connection | φ4mm sockets and flat mini-jack thermocouple socket |
| Pressure connection | Pressure module connector |
| RS232 Interface | standard RS232‐DB9 socket |
| Size | 3.9” x 7.6” x 2.0” (100mm x 192mm x52mm) |
| Weight | 1.6 lb (0.7 kg) |
| Power Supply | Polymer Li‐ion rechargeable battery, or 10V DCadaptor |
| Battery | Rechargeable Li-ion battery (included) |
| Battery Life | 15 hours uninterrupted use Battery life will be reduced when 24V is applied |
| Battery Charge | 110V/220V external power adapter (included) |
Ordering Information
Model number: ConST 317A
Accessories (included)
1pc 110V/220V external power adapter (Available standards: USA, EU)
1pc chargeable Li-ion battery
6pcs test leads
2pcs short circuit cable
|
Model Number
|
Description
|
|
CDP
|
Pressure Modules
|
|
9060
|
Pressure module connection cable
|
|
9050
|
USB to RS 232 (DB-9 Male) Adapter
|
|
9080
|
Cold junction compensation kits
(including TC plug, compensation cable, S, R, B, K ,N, E, J, T)
|
|
9712
|
Spare chargeable Li-ion battery for multifunction calibrators
|
|
9816
|
110V/220V external power adapter (DC 10V)
|
|
9906
|
Carrying case for multifunction calibrators
|
|
9510
|
ConST/CAL Task management software for multifunction calibrators
|